# What changed in FDA Recognition List 066

List 066 (91 FR 54715, 24 August 2026) replaced 28 editions, added 35 recognitions, withdrew 2 and extended one transition period. The rows that matter.

2026-11-10, by Dowelmark.


FDA's Recognition List 066 was published in the Federal Register as FR Doc. 2026-17229 (91 FR 54715), docket FDA-2004-N-0451, and applied from 24 August 2026. It is still the newest list on 10 November 2026. Replayed by Dowelmark it has 94 rows: 28 editions replaced by newer ones, 28 successor editions, 35 new recognitions, 2 withdrawals and 1 change in place. [Every row](/lists/066).

## Replaced by newer editions

The 28 replacements touch 13 specialties. Orthopedic has the most (6), then OB-GYN and Gastroenterology (5, all of ISO 23500 on fluids for haemodialysis) and Cardiovascular (4). A few that many files cite:

- **ISO 10993-1**, biological evaluation: Fifth edition 2018 ([2-258](/recognitions/2-258)) replaced by the Sixth edition 2025 ([2-313](/recognitions/2-313)).
- **ISO 10993-12**, sample preparation: [2-289](/recognitions/2-289) replaced by the Fifth edition with its 2025 amendment ([2-314](/recognitions/2-314)).
- **ISO 20417**, information supplied by the manufacturer: First edition 2021 ([5-135](/recognitions/5-135)) replaced by the Second edition 2026 ([5-149](/recognitions/5-149)).
- **IEC 60601-2-2**, high frequency surgical equipment: Edition 6.0 ([6-389](/recognitions/6-389)) replaced by Edition 6.1 ([6-521](/recognitions/6-521)).
- **ISO 5840-1, -2 and -3**, heart valve prostheses: the 2021 editions replaced ([3-204](/recognitions/3-204), [3-205](/recognitions/3-205), [3-206](/recognitions/3-206)).
- **ISO 10079-1**, suction equipment: [1-157](/recognitions/1-157) replaced by [1-203](/recognitions/1-203).

Replaced is not refused. FDA usually keeps accepting declarations to the old edition for a transition period stated on its database. All 28 replaced editions were still listed in FDA's database on 10 November 2026. [The transition watch](/transition-watch).

## New recognitions

Of the 35 new recognitions, 31 are dental (Dental/ENT, [4-369](/recognitions/4-369) to [4-399](/recognitions/4-399)), from ANSI ADA standards to ISO 9173 and ISO 13397. The others are [ASTM F3766-25](/recognitions/8-650), [IEEE Std 1752.1-2021](/recognitions/13-158), [ISO 7176-31](/recognitions/16-241) and [IEC 80601-2-26](/recognitions/17-23).

## Withdrawn outright

Two standards left the list with no successor: [ASTM F1582-98 (Reapproved 2016)](/recognitions/11-175) and [ISO 8828 Second edition 2014](/recognitions/11-290), both orthopedic.

## Changed in place

[13-142](/recognitions/13-142), FIRST CVSS 3.1, the Common Vulnerability Scoring System: "Transition period extended."

## The database had it first

Every recognition in List 066 entered FDA's database on 25 May 2026, 91 days before the list applied. [Why the database runs ahead](/blog/fda-database-ahead-of-federal-register).

## Hearing about List 067

FDA publishes about two lists a year at no fixed date. Subscribe to the [changes feed](/changes.xml) or call changes_since from an agent with the date of your last check.


## Sources

- [Recognition List 066, 91 FR 54715, FR Doc. 2026-17229, Federal Register, 24 August 2026](https://www.federalregister.gov/documents/2026/08/24/2026-17229/food-and-drug-administration-modernization-act-of-1997-modifications-to-the-list-of-recognized)
- [FDA Recognized Consensus Standards database](https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfStandards/search.cfm)

Drafted with the help of a language model from Dowelmark's own tool output; every figure is a real tool answer.

---

Automated and unreviewed. Dowelmark reads official sources automatically and no person checks each answer. Data can be incomplete, late or wrong. Always verify against the official source linked with every answer before you act. Not legal, regulatory, tax, customs or medical advice. FDA's database and its Federal Register notices decide what FDA recognises, and the official text published by the standards body prevails for the content of any standard. Dowelmark is independent of FDA and of every standards body.

Data version 2026-08-24.list066.db20261110: Federal Register lists to 066 (24 Aug 2026), FDA database read 10 Nov 2026.

Source: Food and Drug Administration, Modifications to the List of Recognized Standards, Recognition Lists published in the Federal Register (federalregister.gov), a US government work; and FDA's Recognized Consensus Standards database. Giorgio Orsucci, the Netherlands. hello@bulkhead.solutions.