Dowelmark
FDA recognition 14-331 ยท Sterility

ISO 11135-1:2007

Sterilization of health care products - Ethylene oxide - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices

Superseded

On 2026-11-29 ISO 11135-1 First edition 2007-05-01 (14-331) is superseded: Recognition List 036 (2014-07-09) replaced it with ISO 11135 Second Edition (14-452); FDA's database no longer lists it.

Answer from recognition_status on 29 November 2026, data version 2026-08-24.list066.db20261110.

Superseded is not refused. FDA usually still accepts declarations of conformity to a replaced edition for a transition period, which it states on the standard's detail page in its database, not in the Federal Register. No longer listed in FDA's database on 2026-11-10, so any transition period has most likely ended. Search FDA's database.
Status on 29 Nov 2026
Superseded
Recognition number
14-331
Specialty
Sterility
Extent
Not listed in FDA's database on 2026-11-10, so no extent is given.
Recognised since
16 Mar 2012, List 028
Ended
9 Jul 2014, List 036
FDA date of entry
Not listed on 10 Nov 2026
In FDA's database
Not listed on 10 Nov 2026
Replaced by
14-452, ISO 11135 Second Edition

Every edition under this line

  1. 14-331 ISO 11135-1 First edition 2007-05-01Superseded, from 16 Mar 2012 to 9 Jul 2014, this page
  2. 14-452 ISO 11135 Second EditionSuperseded, from 9 Jul 2014 to 24 Oct 2019
  3. 14-529 ISO 18472 Second edition 2018-08Superseded, from 24 Oct 2019 to 3 Mar 2021
  4. 14-544 ISO 18472 Second edition 2018-08Recognised, from 3 Mar 2021

What each Federal Register list did

ListApplicableChangeFDA's own wordsCitation
List 02816 Mar 2012RecognisedNo change text: a row in the list's table of new entries77 FR 15765
List 0369 Jul 2014Replaced, by 14-452Withdrawn and replaced with newer version.79 FR 38910

Sources

Automated and unreviewed. Dowelmark reads official sources automatically and no person checks each answer. Data can be incomplete, late or wrong. Always verify against the official source linked with every answer before you act. Not legal, regulatory, tax, customs or medical advice. FDA's database and its Federal Register notices decide what FDA recognises, and the official text published by the standards body prevails for the content of any standard. Dowelmark is independent of FDA and of every standards body.