Dowelmark
FDA recognition 2-141 ยท Biocompatibility

ASTM F1984-99

ASTM F1984-99 (Reapproved 2018) Standard Practice for Testing for Whole Complement Activation in Serum by Solid Materials

Superseded

On 2026-11-29 ASTM F1984-99 (Reapproved 2018) (2-141) is superseded: Recognition List 065 (2026-02-19) replaced it with ASTM F1984-25 (2-308); FDA's database still listed it on 2026-11-10, so it is most likely inside its transition period (end date on the detail page).

Answer from recognition_status on 29 November 2026, data version 2026-08-24.list066.db20261110.

Superseded is not refused. FDA usually still accepts declarations of conformity to a replaced edition for a transition period, which it states on the standard's detail page in its database, not in the Federal Register. Still listed in FDA's database on 2026-11-10, which usually means FDA still accepts declarations of conformity to it inside a transition period. The end date is on its detail page (database.detailUrl); Dowelmark does not hold it yet. Open FDA's detail page.
Status on 29 Nov 2026
Superseded
Recognition number
2-141
Specialty
Biocompatibility
Extent
Complete, from FDA's database
Recognised since
8 Sep 2009, List 022
Ended
19 Feb 2026, List 065
FDA date of entry
14 Jan 2019
In FDA's database
Listed on 10 Nov 2026
Replaced by
2-308, ASTM F1984-25
Replaces
2-95, ASTM F1984-99(2003)

Every edition under this line

  1. 2-95 ASTM F1984-99(2003)Superseded, recognised before 9 Sep 2008 to 8 Sep 2009
  2. 2-141 ASTM F1984-99 (Reapproved 2018)Superseded, from 8 Sep 2009 to 19 Feb 2026, this page
  3. 2-308 ASTM F1984-25Recognised, from 19 Feb 2026

What each Federal Register list did

ListApplicableChangeFDA's own wordsCitation
List 0228 Sep 2009Recognised, replacing 2-95No change text: a row in the list's table of new entries74 FR 46203
List 0235 May 2010Changed in placeExtent of recognition, Relevant guidance and Contact person75 FR 24711
List 03430 Jan 2014Changed in placeExtent of recognition.79 FR 4913
List 0369 Jul 2014Changed in placeReaffirmation.79 FR 38910
List 04426 Jul 2016Changed in placeExtent of recognition, Relevant guidance.81 FR 48809
List 05114 Mar 2019Changed in placeReaffirmation.84 FR 9354
List 06519 Feb 2026Replaced, by 2-308Withdrawn and replaced with newer version.91 FR 7994

Sources

Automated and unreviewed. Dowelmark reads official sources automatically and no person checks each answer. Data can be incomplete, late or wrong. Always verify against the official source linked with every answer before you act. Not legal, regulatory, tax, customs or medical advice. FDA's database and its Federal Register notices decide what FDA recognises, and the official text published by the standards body prevails for the content of any standard. Dowelmark is independent of FDA and of every standards body.