Dowelmark
FDA recognition 2-220 ยท Biocompatibility

ISO 10993-1:2009

ISO 10993-1 Fourth edition 2009-10-15 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process [Including: Technical Corrigendum 1 (2010)]

Superseded

On 2026-11-29 ISO 10993-1 Fourth edition 2009-10-15 (2-220) is superseded: Recognition List 051 (2019-03-14) replaced it with ISO 10993-1 Fifth edition 2018-08 (2-258); FDA's database no longer lists it.

Answer from recognition_status on 29 November 2026, data version 2026-08-24.list066.db20261110.

Superseded is not refused. FDA usually still accepts declarations of conformity to a replaced edition for a transition period, which it states on the standard's detail page in its database, not in the Federal Register. No longer listed in FDA's database on 2026-11-10, so any transition period has most likely ended. Search FDA's database.
Status on 29 Nov 2026
Superseded
Recognition number
2-220
Specialty
Biocompatibility
Extent
Not listed in FDA's database on 2026-11-10, so no extent is given.
Recognised since
14 Aug 2015, List 040
Ended
14 Mar 2019, List 051
FDA date of entry
Not listed on 10 Nov 2026
In FDA's database
Not listed on 10 Nov 2026
Replaced by
2-258, ISO 10993-1 Fifth edition 2018-08
Replaces
2-179, ISO 10993-1 Fourth edition 2009-10-15

Every edition under this line

  1. 2-179 ISO 10993-1 Fourth edition 2009-10-15Superseded, from 16 Mar 2012 to 14 Aug 2015
  2. 2-220 ISO 10993-1 Fourth edition 2009-10-15Superseded, from 14 Aug 2015 to 14 Mar 2019, this page
  3. 2-258 ISO 10993-1 Fifth edition 2018-08Superseded, from 14 Mar 2019 to 24 Aug 2026
  4. 2-313 ISO 10993-1 Sixth edition 2025-11Recognised, from 24 Aug 2026

What each Federal Register list did

ListApplicableChangeFDA's own wordsCitation
List 04014 Aug 2015Recognised, replacing 2-179No change text: a row in the list's table of new entries80 FR 48871
List 04426 Jul 2016Changed in placeExtent of recognition, Relevant guidance.81 FR 48809
List 05114 Mar 2019Replaced, by 2-258Withdrawn and replaced with newer version.84 FR 9354

Sources

Automated and unreviewed. Dowelmark reads official sources automatically and no person checks each answer. Data can be incomplete, late or wrong. Always verify against the official source linked with every answer before you act. Not legal, regulatory, tax, customs or medical advice. FDA's database and its Federal Register notices decide what FDA recognises, and the official text published by the standards body prevails for the content of any standard. Dowelmark is independent of FDA and of every standards body.