Dowelmark
FDA recognition 3-67 ยท Cardiovascular

ASTM F2129-06

ASTM F2129-06 Standard Test Method for Conducting Cyclic Potentiodynamic Polarization Measurements to Determine the Corrosion Susceptibility of Small Implant Devices

Superseded

On 2026-11-29 ASTM F2129-06 (3-67) is superseded: Recognition List 023 (2010-05-05) replaced it with ASTM F2129-08 (3-76); FDA's database no longer lists it.

Answer from recognition_status on 29 November 2026, data version 2026-08-24.list066.db20261110.

Superseded is not refused. FDA usually still accepts declarations of conformity to a replaced edition for a transition period, which it states on the standard's detail page in its database, not in the Federal Register. No longer listed in FDA's database on 2026-11-10, so any transition period has most likely ended. Search FDA's database.
Status on 29 Nov 2026
Superseded
Recognition number
3-67
Specialty
Cardiovascular
Extent
Not listed in FDA's database on 2026-11-10, so no extent is given.
Recognised since
9 Sep 2008, List 020
Ended
5 May 2010, List 023
FDA date of entry
Not listed on 10 Nov 2026
In FDA's database
Not listed on 10 Nov 2026
Replaced by
3-76, ASTM F2129-08
Replaces
3-57,

Every edition under this line

  1. 3-57 Citation not printed in the listsSuperseded, recognised before 9 Sep 2008 to 9 Sep 2008
  2. 3-67 ASTM F2129-06Superseded, from 9 Sep 2008 to 5 May 2010, this page
  3. 3-76 ASTM F2129-08Withdrawn, from 5 May 2010 to 14 Aug 2015

What each Federal Register list did

ListApplicableChangeFDA's own wordsCitation
List 0209 Sep 2008Recognised, replacing 3-57No change text: a row in the list's table of new entries73 FR 52358
List 02118 Mar 2009Changed in placeOffices, Devices Affected, Relevant Guidance, CFR Citation and Product Codes, Contact Person74 FR 11586
List 02118 Mar 2009Changed in placeOffices, Devices affected, Type of Standard, Product code, Relevant guidance, Contact person74 FR 11586
List 0235 May 2010Replaced, by 3-76Withdrawn and replaced with newer version75 FR 24711

Sources

Automated and unreviewed. Dowelmark reads official sources automatically and no person checks each answer. Data can be incomplete, late or wrong. Always verify against the official source linked with every answer before you act. Not legal, regulatory, tax, customs or medical advice. FDA's database and its Federal Register notices decide what FDA recognises, and the official text published by the standards body prevails for the content of any standard. Dowelmark is independent of FDA and of every standards body.