Dowelmark
FDA recognition 6-389 ยท General Hospital/General Plastic Surgery

IEC 60601-2-2:2017

IEC 60601-2-2 Edition 6.0 2017-03 Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories.

Superseded

On 2026-11-29 IEC 60601-2-2 Edition 6.0 2017-03 (6-389) is superseded: Recognition List 066 (2026-08-24) replaced it with IEC 60601-2-2 Edition 6.1 2023-02 CONSOLIDATED VERSION Corrected Version 2025-11 (6-521); FDA's database still listed it on 2026-11-10, so it is most likely inside its transition period (end date on the detail page).

Answer from recognition_status on 29 November 2026, data version 2026-08-24.list066.db20261110.

Superseded is not refused. FDA usually still accepts declarations of conformity to a replaced edition for a transition period, which it states on the standard's detail page in its database, not in the Federal Register. Still listed in FDA's database on 2026-11-10, which usually means FDA still accepts declarations of conformity to it inside a transition period. The end date is on its detail page (database.detailUrl); Dowelmark does not hold it yet. Open FDA's detail page.
Status on 29 Nov 2026
Superseded
Recognition number
6-389
Specialty
General Hospital/General Plastic Surgery
Extent
Partial, from FDA's database
Recognised since
21 Aug 2017, List 047
Ended
24 Aug 2026, List 066
FDA date of entry
7 Jun 2021
In FDA's database
Listed on 10 Nov 2026
Replaced by
6-521, IEC 60601-2-2 Edition 6.1 2023-02 CONSOLIDATED VERSION Corrected Version 2025-11
Replaces
6-336, IEC 60601-2-2 Edition 5.0 2009-02

Every edition under this line

  1. 6-197 IEC 60601-2-2 Ed. 1.0Superseded, recognised before 10 Jun 2010 to 10 Jun 2010
  2. 6-228 IEC 60601-2-2 Edition 5.0Superseded, from 10 Jun 2010 to 27 Jan 2015
  3. 6-336 IEC 60601-2-2 Edition 5.0 2009-02Superseded, from 27 Jan 2015 to 21 Aug 2017
  4. 6-389 IEC 60601-2-2 Edition 6.0 2017-03Superseded, from 21 Aug 2017 to 24 Aug 2026, this page
  5. 6-521 IEC 60601-2-2 Edition 6.1 2023-02 CONSOLIDATED VERSION Corrected Version 2025-11Recognised, from 24 Aug 2026

What each Federal Register list did

ListApplicableChangeFDA's own wordsCitation
List 04721 Aug 2017Recognised, replacing 6-336No change text: a row in the list's table of new entries82 FR 39591
List 0569 Dec 2021Changed in placeExtent of recognition.86 FR 70132
List 06624 Aug 2026Replaced, by 6-521Extent of recognition. Withdrawn and replaced with newer version.91 FR 54715

Sources

Automated and unreviewed. Dowelmark reads official sources automatically and no person checks each answer. Data can be incomplete, late or wrong. Always verify against the official source linked with every answer before you act. Not legal, regulatory, tax, customs or medical advice. FDA's database and its Federal Register notices decide what FDA recognises, and the official text published by the standards body prevails for the content of any standard. Dowelmark is independent of FDA and of every standards body.