Dowelmark
FDA recognition 8-190 ยท Materials

ASTM F90-09

ASTM F 90-09 Standard Specification for Wrought Cobalt-20 Chromium-15 Tungsten-10 Nickel Alloy for Surgical Implant Applications (UNS R30605)

Superseded

On 2026-11-29 ASTM F90-09 (8-190) is superseded: Recognition List 041 (2016-04-04) replaced it with ASTM F90-14 (8-399); FDA's database no longer lists it.

Answer from recognition_status on 29 November 2026, data version 2026-08-24.list066.db20261110.

Superseded is not refused. FDA usually still accepts declarations of conformity to a replaced edition for a transition period, which it states on the standard's detail page in its database, not in the Federal Register. No longer listed in FDA's database on 2026-11-10, so any transition period has most likely ended. Search FDA's database.
Status on 29 Nov 2026
Superseded
Recognition number
8-190
Specialty
Materials
Extent
Not listed in FDA's database on 2026-11-10, so no extent is given.
Recognised since
5 May 2010, List 023
Ended
4 Apr 2016, List 041
FDA date of entry
Not listed on 10 Nov 2026
In FDA's database
Not listed on 10 Nov 2026
Replaced by
8-399, ASTM F90-14
Replaces
8-145, ASTM F0090-07

Every edition under this line

  1. 8-145 ASTM F0090-07Superseded, recognised before 9 Sep 2008 to 5 May 2010
  2. 8-190 ASTM F90-09Superseded, from 5 May 2010 to 4 Apr 2016, this page
  3. 8-399 ASTM F90-14Superseded, from 4 Apr 2016 to 19 Feb 2026
  4. 8-630 ASTM F90-24Recognised, from 19 Feb 2026

What each Federal Register list did

ListApplicableChangeFDA's own wordsCitation
List 0235 May 2010Recognised, replacing 8-145No change text: a row in the list's table of new entries75 FR 24711
List 03430 Jan 2014Changed in placeExtent of recognition.79 FR 4913
List 0414 Apr 2016Replaced, by 8-399Withdrawn and replaced with newer version.81 FR 19196

Sources

Automated and unreviewed. Dowelmark reads official sources automatically and no person checks each answer. Data can be incomplete, late or wrong. Always verify against the official source linked with every answer before you act. Not legal, regulatory, tax, customs or medical advice. FDA's database and its Federal Register notices decide what FDA recognises, and the official text published by the standards body prevails for the content of any standard. Dowelmark is independent of FDA and of every standards body.